Bio-chemistry Analyser
- Home
- Pathology Instruments
- Bio-chemistry Analyser
Product Code: KI-19-519
| Labtech Biochemistry Analyzer for routine chemistry , enzymes & drug levels . It has a user friendly on board software . It is very economical , reliable high precision machine. Featues:
|
|
| Specifications: | |
| Measuring System | Photometric |
| Linear Measurement Range | 0.000 to 3.0000 AbsorbtionUnits (A) |
| Photometric Accuracy | ?2% or 0.007 whichever is higher.from 0 to 1.5 A , ?3% from 1.5A to 3.0A |
| Drift | <0.007 A/hr |
| Photometric Linearity | 2.2A |
| Optical Measurement | Photodiode |
| Filers: | |
| Type | Interference Filter |
| Wavelength | 340,405,510,546, 578 & 630 nm & Two Optional |
| Half Bandwidth | 10 nm ? 2nm |
| Selection | Automatic by Stepper Motor |
| ? | ? |
| Cuvette Volume | 500 ?l |
| Temperature of Cuvette Block | 370C, Peltier Controlled |
| Light Source | xenon gas filled / Tungsten Halogen |
| Warm upTime | 90 Sec |
| Display | Four Line LCD , backlit , 16 characters |
| Printer Interface | Optional Parallel Report |
| Memory | 8 KB Non-Volatile RAM , Battery back up supports 40 open tests |
| Analysis Mode | Absorbance : Fixed Time |
| ? | End Point : Kinetic |
| ? | Multi -Standard :Differential |
| ? | Ratio : PTT |
| Concentration Calculation | By factor or By Standard |
| RS232 Serial Port | 2400 baud , 1 start , 8 data , 1 stop ,no parity bits |
| Power | Wattage: 50 watts |
| ? | Voltage : 115-230 Volts ? 10% , 50/60 HZ |
| Operating Positions | on horizontal flat , rigid & vibration Free space |
| Operating Conditions | Temperature: From + 18oC to 35oC |
| ? | Relative Humidity: up to 85% |
| Storage Conditions | Temperature: From -10oC to 60oC |
| ? | Relative Humidity: up to 85% |
| Enclosures | Abs Fire Retardant |
| Size (cms) | 30 x 38 x 13.5 (lxbxh) |
| Weight | 5 Kgs |