Enhanced safety – A topmost priorty S-ORC (Sensitive Oxygen Ratio Controller)
The S-ORC ensures a minimum fresh gas oxygen concentration of 25% at all times to avoid hypoxia. The N2O is cut off automatically if the Oxygen supply fails or when the O2 flow meter knob is inadvertently closed. The patient can never get ventilated on N2O alone.
Many key components are imported to insure highest level of reliability.
Fitted with special, second stage regulator for stable working pressure and safety.
Circle System ISO 9
International standard circle system with highest absorption efficiency. Easy to refill & change canisters. Unique designs allows for use of single or dual canisters. Total absorber capacity of
2.5 Ltr.
Incorporates calibrates Airway Pressure Limiting Valve (APL) and single outlet for waste gas.
Circle absorber can be connected to gas scavenging system.
Read Technical Data
? | ? | ? |
Pipeline Inlet Pressure | O2 N2 O Air | 2.7 – 6 Bar 2.7 – 6 Bar 2.7 – 6 Bar |
Cylinder Inlet | Four Pin Index Yokes Two each for Oxygen and Nitrous Oxide Cylinders | |
Cylinder pressure reducer | Individual pressure reducer for each yoke followed by a common second stage for stable working pressure. | |
Flowmeter range | O2 N2 O Air | 0 – 10 Ltrs. / min. 0 – 12 Ltrs. / min. 0 – 12 Ltrs. / min. ( Optional ) |
Safety | ORC (Optional) | ?Avishkar Oxygen Ratio Controller ensures minimum 25% Oxygen at all the time. Model – KI-043. |
Nitrolock | Basic safety against accidental delivery of Nitrous Oxide, when Oxygen flowmeter is closed. Model – KI-043 (NL). | |
Oxyfailure Alarm | Audible alarm for minimum seven seconds in case of Oxygen supply pressure drops below 2 kPa X 100. | |
pressure relief valve | Mechanical safety against increase in low pressure circuit set at 20 kPa. | |
Oxygen Flush | Up to 55 Ltrs. of Oxygen flow, with spring-loaded push switch. | |
Changeover unit | Toggle switch for changing from open to close circuit and vice versa | |
Breathing Circuit | Magill’s Breathing Circuit – Complete with angle mount, Breathing Hose, Reservior bag, Expiratory valve and face mask size 4. | |
Dimensions | Overall Top tray Overall | 810 mm X 860 mm X 1450 mm 610 mm X 3000 mm X 15 mm 65 Kg. |
Models | KI-043 (NL) KI-043 (ORC) KI-043 (ISI). As per IS 11378 ( Equivalent to ISO 5358 ) Available in both NL & ORC models. | |
Optional Accessories | Vapour 2000 for Halothane, Isoflurane or Sevoflurane Anaesthesia Ventilator |